@ -59,14 +59,6 @@ The SuperSensor's voice functionality can be completely disabled if voice
|
||||
support is not desired. This defeats most of the point of the SuperSensor,
|
||||
but can be done if desired.
|
||||
|
||||
### Gas Ceiling
|
||||
|
||||
The AQ (air quality) calculation from the BME680 requires a "maximum"/ceiling
|
||||
threshold for the gas resistance value in clean air after some operation
|
||||
time. The value defaults to 200 kΩ to provide an initial baseline, but
|
||||
should be calibrated manually after setup as each sensor is different. See
|
||||
the section "Calibrating AQ" below for more details.
|
||||
|
||||
### Light Threshold Control
|
||||
|
||||
The SuperSensor features a "light presence" binary sensor based on the light
|
||||
@ -178,139 +170,3 @@ For detect, no occupancy will ever fire.
|
||||
For clear, no states will clear occupancy; with any detect option, this
|
||||
means that occupancy will be detected only once and never clear, which
|
||||
is likely not useful.
|
||||
|
||||
## Calibrating AQ
|
||||
|
||||
The Supersensor uses the Bosch BME680 combination temperature, humidity,
|
||||
pressure, and gas sensor to provide a wide range of useful information about
|
||||
the environmental conditions the sensor is placed in. However, this sensor
|
||||
can be tricky to work with.
|
||||
|
||||
While it's normally recommended to use the Bosch BSEC library with this
|
||||
sensor, in my ~6 month experience I found this library to be far more trouble
|
||||
than it was worth. Specifically, it's IAQ measurement is nearly useless, with
|
||||
a strong tendency to get stuck in an upward trend constantly "calibrating"
|
||||
itself to higher and higher baselines, to the point where nonsensical values
|
||||
were being read. After much research into this, I decided to abandon the
|
||||
library in version 1.1 and went with a more custom solution.
|
||||
|
||||
Instead of the BSEC, we use the stock BME680 ESPHome library, along with
|
||||
some calculations by thstielow on GitHub in their [IAQ project](https://github.com/thstielow/raspi-bme680-iaq).
|
||||
This provided some useful example code and formulae to calculate a useful
|
||||
Air Quality (AQ) value instead of the useless Bosch value.
|
||||
|
||||
However using this method requires some manual calibration of the sensor
|
||||
after putting it together but before final use, in order to get a somewhat
|
||||
accurate value out of the AQ component. If you don't care about the AQ value,
|
||||
you can skip this, but it is recommended to take full advantage of the sensor.
|
||||
|
||||
As a quick explainer, the code leverages a combination of the "Gas Resistance"
|
||||
value provided by the sensor, along with an absolute humidity calculated from
|
||||
the temperature and relative humidity of the sensor (included ESPHome sensor),
|
||||
along with two values (one configurable, one hard-coded) and several formulae
|
||||
to arrive at the resulting AQ value. For full details of the calculation,
|
||||
see the repository linked above, which was re-implemented faithfully here.
|
||||
|
||||
The first thing to note is that each BME680 sensor is wildly different in
|
||||
terms of gas resistance values. In the same air, I had sensors reading values
|
||||
that differed by nearly 200,000Ω, which necessitates a human-configurable
|
||||
baseline value. Further, the IAQ project recommends determining a linear
|
||||
slope value for this, but instead of trying to explain how to calculate this,
|
||||
I just went with the default slope value of 0.03 for this first iteration.
|
||||
|
||||
Thus, the main difficulty in getting a useful AQ score is finding the
|
||||
"Gas Resistance Ceiling" value. This value is configurable in the
|
||||
SuperSensor interface (Web or HomeAssistant), and should be calibrated as
|
||||
follows during the initial setup of the supersensor.
|
||||
|
||||
1. Find a known-clean room, for instance a well-ventilated, well-cleaned
|
||||
room in your house or similar. It should have fresh air (no stray VOCs) but
|
||||
also minimal drafts or outside exposure especially if there is a poor external
|
||||
AQ level. This will be your calibration reference room. Ideally, this room
|
||||
should be somewhere between 16C and 26C for optimal performance, so air
|
||||
conditioning (or a nice spring/fall day) is best.
|
||||
|
||||
2. Turn on the SuperSensor in this environment, and connect it to your
|
||||
HomeAssistant instance; this will be critical for viewing historical graphs
|
||||
during the following steps.
|
||||
|
||||
3. Let the SuperSensor run to "burn in" the gas sensor for at least 3-6 hours,
|
||||
or until the value for the Gas Resistance stabilizes. It is best to avoid much
|
||||
movement or activity in the selected calibration room to avoid disrupting
|
||||
the sensor during this time. It is also best to ensure that the ambient
|
||||
temperature changes as little as possible during this time.
|
||||
|
||||
4. Review the resulting graph of Gas Resistance over the burn-in period. You
|
||||
can usually ignore the first hour or two as the sensor was burning in, and
|
||||
focus instead on the last hour or so.
|
||||
|
||||
5. Make note of the highest mean value reached by the sensor during this time.
|
||||
This will be your baseline value for calibrating the Gas Resistance Ceiling.
|
||||
|
||||
6. Round the value up to the nearest 1000. For example, if the maximum value
|
||||
was 195732.1, round this to 196000.0.
|
||||
|
||||
7. Find the difference in the temperature of the BME680 temperature sensor
|
||||
from 20C, called ΔT below. I found this part by trial-and-error, so this is
|
||||
not precise, but as an example if the calibration room is reporting 26C, your
|
||||
ΔT value in the next step is 6. If your temperature was below 20C, use 0.
|
||||
|
||||
8. Use one of the following formulae to come up with your offset value, which
|
||||
depends on the maximum value range found in step 6.
|
||||
|
||||
* `<100,000`: 200 * ΔT = 0-1200
|
||||
* `100,000-200,000`: 500 * ΔT = 0-3000
|
||||
* `>200,000`: 1000 * ΔT = 0-6000
|
||||
|
||||
Again this value is rough, and might not even really be needed, but helps
|
||||
avoid weird issues with AQ values dropping suddenly later as temperature
|
||||
and humidity changes.
|
||||
|
||||
9. Add your offset value from step 8 to the rounded maximum from step 6.
|
||||
For example, 196000.0 with a ΔT of 5C (25C ambient) yields 201000.0
|
||||
|
||||
10. Divide the result from 9 by 1000 to give a number from 1-500. This
|
||||
is the value to enter as the "Gas Resistance Ceiling (kΩ)" for this
|
||||
sensor. This value will be saved in the NV-RAM of the ESP32 and preserved
|
||||
on reboots.
|
||||
|
||||
At this point, you should have a value that results in the "BME680 AQ"
|
||||
sensor reporting 100% AQ, i.e. clean air. You can now test to ensure
|
||||
that the value will correctly drop as VOCs are added.
|
||||
|
||||
1. Take a Sharpie permanent marker, Acetone nail polish remover, or some
|
||||
other VOC that the BME680 gas sensor can detect, and place it near the
|
||||
sensor. For example with a sharpie, remove the cap and place the tip
|
||||
about 1-2cm from the sensor, or place a small capful of nail polish
|
||||
remover about 3-5cm from the sensor.
|
||||
|
||||
2. Wait about 30 seconds.
|
||||
|
||||
3. You should see the AQ value drop precipitously, into the order of 50%
|
||||
or lower, and ideally closer to 0-20%. If the value remains higher than
|
||||
50% with this test, your calculated Gas Resistance Ceiling might be
|
||||
too low, and should be increased in increments of 1000.
|
||||
|
||||
4. Remove the VOC source (replace the cap, remove the capful of remover,
|
||||
etc.) and wait about 30-60 minutes.
|
||||
|
||||
5. You should see the AQ value and gas resistance return to their original
|
||||
values. If it is significantly lower than before, even after waiting 60+
|
||||
minutes, restart the calculation from step 5 in the previous section
|
||||
using this new value as the baseline.
|
||||
|
||||
At this point, the sensor should be calibrated enough for day-to-day
|
||||
casual home use, and will tell you if there is any significant
|
||||
VOC contamination in the air by dropping the AQ value from 100% to some
|
||||
lower value representing the approximate decrease in air quality. Since
|
||||
the sensor also factors in the absolute humidity (and via that, the
|
||||
ambient temperature) into the AQ calculation, high humidity will also
|
||||
drop the value, as this too impacts the air quality. Hopefully this
|
||||
is useful for your purposes.
|
||||
|
||||
If you find that the AQ value still doesn't represent known reality,
|
||||
you can also tweak the in-code value for `ph_slope` on line 522, as
|
||||
it's possible your sensor differs significantly here. As mentioned
|
||||
above this is still a work in progress to determine for myself, so
|
||||
future versions may alter this or include calibration of this value
|
||||
automatically, depending on how things go in my testing.
|
||||
|